Chapter 2 Dispensing to SpectroCHIP Arrays
32
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
Loading a method
A method file specifies dispensing run settings, such as the mapping file, pin cleaning
operations to perform, and speed at which the pin array moves.
The instrument includes predefined method files for some common types of dispensing.
You can also create custom method files. See
for more details on
creating custom method files.
To load a method file:
1.
Go to the Main Menu (tap the
BACK
button
on any screen).
2.
Tap
TRANSFER o
n the Main Menu.
NOTE
NOTE
Do not tap the OPEN button
on the Transfer screen as it will cause a calibrant velocity error.
3.
On the
Transfer
screen, tap
METHODS
, then tap
OPEN
on the Method
screen.
4.
Select a method file in the list that appears by tapping the file name
Figure 2.14 Main Menu and Transfer Screen
Figure 2.15 Transfer Screen and Method Screen