44
Power Quality Analyzer Model 3945
5.4 Transient
Mode
Press the transient display mode button -
Transients are displayed in the form of waveforms. All channels (up to 6, based
upon confi guration) are stored in memory for each transient. Up to 50 transients
can be captured.
The function buttons have the following functions in this mode:
- Starts capture search programming for
a new transient
- Stops capture search
- Displays a captured transient
- Deletes a captured transient
07/25/02 10:54
100%
SEARCH FOR NEW TRANSIENTS
SLOTS AVAILABLE
START
: 07/25/02 10:55
50
: 07/25/02 10:55
: 1 %
: 1 %
: 1
: T E S T
END
V threshold
A threshold
Number
NAME
1
2
3
Figure 5-15
Transient recording start and end time.
Trigger threshold: selection of 1%, 2%, 5%, 10%, 20%, 50%, 100% for voltage
and current are available.
- Press the
buttons to select either “
V
” or “
A
”, then use the
buttons
to modify the trigger threshold.
Choose a name for the session and the number of transients to be captured
with the arrow buttons:
-
: Selection of the character place (7 characters max)
-
: Selection of the alphanumeric value
-
: Press the Enter button to apply the conditions and to activate
transient capture