16
GroundFlex™ Adapter Model 6474
For test frequencies between 41 and 5087 Hz, between 1 and 4 turns of the
GroundFlex
™
sensor, and a selection of 1 to 4 measurement channels, the
operating accuracies are the following:
SENSITIVITY and minimum I
SEL
minimal
Operating accuracy for R
S-ES
S x 1/10
I
SEL
> 10mA
± (10% + 4cts)
S x 1
I
SEL
> 5mA
± (5% + 4cts)
S x 10
I
SEL
> 5mA
± (5% + 4cts)
5mA > I
SEL
> 0.5mA
± (15% + 10cts)
Current Measurement with GroundFlex™
Measurement Range
0.0 to
99.9mA
(1)
100 to
999mA
1.00 to
9.99A
10.0 to
99.9A
Resolution
0.1mA
(1)
1mA
0.01A
0.1A
Frequency
16 to 400Hz
Accuracy
<5% of Reading @ 50/60Hz
(1): Valid only in the SENSITIVITY x 10 setting
The current measured depends on the number of turns of the GroundFlex
™
sensor
around the conductor to be measured. (e.g. i
f the
GroundFlex
™
sensor is wrapped
4 times around a conductor carrying a current of 1A, the input signal will be the
same as if the
GroundFlex
™
sensor were wrapped once around a conductor car-
rying a current of 4A. The operating error is therefore indicated for an input signal
current in A.tr (Ampere.turns)
.
The minimum current measured depends on the setting of the SENSITIVITY switch
as indicated by the table below:
SENSITIVITY
I
MIN
(A.tr)
Points additional error (pt-er)
x 10
0.01
5
x 1
0.04
2
x 1/10
0.16
2
Accuracy
Current (A * turn)
16 to 49Hz
50 to 99Hz
100 to 400Hz
I
MIN
to 0.399
± (20% + pt-er)
± (5% + pt-er)
± (3% + pt-er)
0.4 to 39.9
± (10% + 2ct)
± (3% + 2ct)
± (3% + 2ct)
40 to 99.9
± (10% + 2ct)
± (3% + 2ct)
± (20% + 2ct)