GroundFlex™ Adapter Model 6474
9
1.
Set the switch of the 6472 to
SET-UP
.
2.
Set the
SENSOR TURNS
switch to 4.
3.
Set the
SENSITIVITY
switch to x1.
4.
Set the
INPUT SELECTION
switch to 1.
1-2-3-4
1-2-3
1-2
1
2
3
4
SENSOR TURNS
SENSITIVITY
INPUT
5.
Press the
Hz/OPTIONS
button 5 times.
The device starts by comparing the current I
Int
injected by the device into the
calibration loop to the current I
A
measured by the GroundFlex
™
sensor. If they
are different, the "CAL" symbol blinks to recommend a calibration.
x 5
4
H
E
S x 1
1
AUTO
mA
m A
NOTE:
To avoid measurement errors, do not touch the GroundFlex
™
sensor or
the calibration loop during the calibration.
6.
To start the calibration, press
2nd
+
START
.