12-17-2018
Information subject to change
9
3110 HELP FILES – SECTION 3
Figure 3.2 – System Calibration tab.
Figure 3.3 – System Calibration settings.
System Calibration
(continued on next page)
Figure 3.4 – Running the Calibration Test.
1. Select the Settings but-
ton and then open the
System Calibration tab.
2. Adjust the 3110 Output Volt-
age, if desired. A default test
signal of 1 Vp will be pre-
entered in this box. Note: the
default DC offset is 0 and
cannot be changed.
3. Connect an oscilloscope to
the DUT (load at the system
output).
4. Press the Run Calibration
Test button to begin the Cali-
bration Test. The 3110 will
generate a 1 kHz sine wave
signal at the designated test
voltage.