Fig. 1. Block diagram of Alpha Spectrometer.
ORTEC MODEL A-576 A-PAD
Preamplifier, Amplifier, Discriminator with Bias Supply
1. DESCRIPTION
The ORTEC Model A-576 A-PAD is a single- wide
The energy ranges controllable from the front panel
NIM module intended for performing alpha
are 3 to 8 MeV, 4 to 7 MeV, 3 to 5 MeV, 4 to 6 MeV, 5
spectroscopy and counting when used in conjunction
to 7 MeV, and 6 to 8 MeV. The internal test pulser can
with a silicon surface barrier detector. The instrument
provide a signal at 5 MeV, 6 MeV, or 7 MeV by
includes a variable detector bias supply, a
internal jumper selection.
preamplifier, a shaping and stretching amplifier, a
biased amplifier, a test pulser, and a discriminator
In addition to the calibrated Energy output, the
(Fig. 1). The detector bias supply and the amplifier
COUNT front-panel BNC connector provides a NIM-
have selectable polarity so that it can be ued with any
standard positive logic pulse for gross alpha counting
charged-particle silicon semiconductor detector.
and/or routing. The DET BIAS MONITOR front-panel
Examples are ORTEC's conventional and
test points permit monitoring detector leakage current
Ruggedized
-
Si surface barrier detectors and its ion-
and detector bias voltage.
implanted Si detectors.