background image

6

5.  THEORY OF OPERATION

The complete schematic for the A-576 A-PAD (No.

amplitude for 5, 6, and 7 MeV to the input to the

752500)  is included  at the back  of this manual.   

preamplifier. A screwdriver calibration, included on the

Fig. 5 is a block diagram of the electronics.

PWB for factory adjustment of the test pulser circuit,

The detector voltage (TEST/BIAS/OFF) switch on the

front-panel toggle switch is set at BIAS, the test pulser

front panel turns power ON and OFF for the variable

is turned off and does not appear in the output

100-V bias power supply. The power is on to the BIAS

spectrum.

supply in both the TEST and BIAS positions.

The polarity is set by Jumper J1, which is actually

inverted by the preamplifier and are furnished as the

three separate jumpers (see schematic). The indicated

input to the amplifier that includes the first integration

polarity is negative, which is appropriate when an

and differentiation shaping circuits. The amplifier can

ORTEC Ruggedized detector is used. When the

be calibrated by a factory-adjusted trim potentiometer

detector voltage is negative, its output pulses are

on the PWB. The output is a shaped negative pulse

positive. When they pass through the inverting charge-

that can be observed at test point TP3. 

sensitive preamplifier, they are negative, and this is
the polarity that is then selected by the pair of jumpers

The pulse at TP3 is furnished into a fixed-level

(J3) on the PWB. If a conventional surface barrier, or

discriminator (U3), and through a diode dc-restorer

ion-implanted, detector is used, all five of these

and integration network. If the amplitude represents

jumpers must be changed to "+" because these

~2.5 MeV or more, the discriminator fires and

detectors, positively biased, generate negative output

generates a positive-shaped pulse that is used to

pulses, which are then inverted, resulting in positive

enable the stretcher. This pulse is also provided as the

pulses at the J3 location in the circuit.

COUNT output signal. The pulse from the second

The 100-V power supply output on the PWB can be

If the stretcher is enabled, the peak amplitude is

tested for both polarity and amplitude at test point

stretched to improve the measurement accuracy of the

TP1, which is identified on the PWB. The variable 0 to

multichannel analyzer, which uses the ENERGY

100 V bias voltage can be measured at the front-panel

output for measurement. If the stretcher is not enabled

test point jack labeled HV. 

(because the pulse represents <2.5 MeV), the output

The detector leakage current can be measured across

stretched. The output of the stretcher is a negative

the 1.1-M

6

 resistor between front-panel test points HV

pulse that can be monitored at test point TP4 on 

and DET, using a 10-M

6

 impedance meter. The

the PWB.

voltage (in volts) converts to µA of detector current.
The output of the Bias supply is applied  through a

The biased amplifier accepts the stretcher output and

high impedance (totaling 11 M

6

) to the dc connection

the bias level selected by the front-panel ENERGY

between the detector and the preamplifier.

RANGE switch (low end of range). The bias level, or

The output from the charge-sensitive preamplifier can

at 3, 4, 5, or 6 MeV by the ENERGY RANGE switch,

be checked at TP2.  The nominal conversion gain is

which can be adjusted by ±10% by the front-panel

45 mV/MeV. The preamplifier gain from the External

screwdriver control, E BIAS. 

Pulser input is nominally "

1" at TP2. 

The pulser switch (TEST/BIAS/OFF) on the front

bias level from the input pulse amplitude and to

panel turns power on and off for the test pulser. The

amplify the excess amplitude by a factor of 10 on the

test pulser can be set to furnish the equivalent pulse

3

&

5, 4

&

6, 5

&

7, and 6

&

8 MeV ranges, and by a factor 

should not be changed by the customer. When the

Pulses from the detector and/or the test pulser are

integration shaping circuit is furnished to the stretcher.

is passed to the biased amplifier without being

lower level, accepted into the bias amplifier can be set

The function of the biased amplifier is to subtract the

Содержание ORTEC A-576 A-PAD

Страница 1: ...Model A 576 A PAD Preamplifier Amplifier Discriminator with Bias Supply Operating and Service Manual Printed in U S A ORTEC Part No 761930 0202 Manual Revision D ...

Страница 2: ... that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the fault of the instrument being returned and of the model serial and revision Rev on rear panel numbers Failure to do so may cause unnecessary delays in getting the unit repaired The ORTEC standard proce...

Страница 3: ...NECTION TO POWER 4 3 4 OUTPUT CONNECTIONS 4 3 5 ENERGY OUTPUT FULL SCALE ADJUSTMENT 4 4 OPERATION 5 4 1 GENERAL 5 4 2 CONNECTING A DETECTOR 5 5 THEORY OF OPERATION 6 LIST OF FIGURES Fig 1 Block diagram of Alpha Spectrometer 1 Fig 2 Jumper Locations for Setting Polarity on the PWB 3 Fig 3 Side Panel Test Pulser Selection 4 Fig 4 Typical A PAD Output Waveforms with 5 MeV Test Pulser 5 Fig 5 Detailed...

Страница 4: ...iv ...

Страница 5: ...MeV by includes a variable detector bias supply a internal jumper selection preamplifier a shaping and stretching amplifier a biased amplifier a test pulser and a discriminator In addition to the calibrated Energy output the Fig 1 The detector bias supply and the amplifier COUNT front panel BNC connector provides a NIM have selectable polarity so that it can be ued with any standard positive logic...

Страница 6: ...o give a 4 µs wide positive output pulse INPUT BIAS POLARITY PWB jumpers allow ENERGY RANGES 3 TO 8 MeV 4 to 7 MeV 3 to 5 polarity Negative for ORTEC Surface Barrier or Ion MeV 4 to 6 MeV 5 to 7 MeV 6 to 8 MeV Implanted Ruggedized detectors or positive for conventional detectors Shipped in Negative position CALIBRATED TEST PULSER Equivalent Energy 5 MeV 6 MeV or 7 MeV Shipped at 5 MeV Rate is 100 ...

Страница 7: ...are accessible when the side panel is removed from the instrument chassis Five of the jumpers are used to select either positive or negative detector voltage polarity and the corresponding correct amplifier polarity Negative bias is required for ORTEC Ruggedized detectors positive for conventional surface barrier or ion implanted detectors These five internal jumpers need to be changed only when a...

Страница 8: ...ut PREAMP IN connector on the A PAD For best Continue with operation of the A 576 results keep the length of this cable as short as possible due to the capacitive loading of the cable to minimize preamplifier noise and maintain the preamplifier stability Type RG 62 U cable is recommended for the detector to preamplifier connection This is 93 6 cable with a capacity of 13 5 pF ft 3 3 CONNECTION TO ...

Страница 9: ...e bias voltage with a 6 µs width shown in Fig 4 ON may damage the FET transistor at the input of the preamplifier or destroy the detector 4 System electronic noise can be tested by Then proceed as follows the MCA from the ENERGY output 1 With the bias OFF connect the detector to the 5 The test pulser can then be turned OFF by moving A PAD with a short cable preferably RG 62A U the TEST BIAS OFF sw...

Страница 10: ...es at the J3 location in the circuit COUNT output signal The pulse from the second The 100 V power supply output on the PWB can be If the stretcher is enabled the peak amplitude is tested for both polarity and amplitude at test point stretched to improve the measurement accuracy of the TP1 which is identified on the PWB The variable 0 to multichannel analyzer which uses the ENERGY 100 V bias volta...

Страница 11: ...etected input energy exceeds the selected bias MeV range The output of the biased amplifier is a level The full scale output of 10 V can be adjusted positive pulse with an amplitude in the range from 0 1 down to 7 75 V using the front panel screwdriver to 10 V which is proportional to the amount by which control E ...

Отзывы: