![Advanced Instruments 3320 Скачать руководство пользователя страница 52](http://html.mh-extra.com/html/advanced-instruments/3320/3320_user-manual_2856031052.webp)
Advanced
®
Micro-Osmometer Model 3320 User’s Guide
Maintaining the instrument
1. Cleaning Air Vents
Examine the air vents on the underside and rear of the instrument to
ensure that they are unobstructed by dust or debris.
2. Cleaning the Instrument Exterior
Use a mixture of warm water and soap. Isopropyl Alcohol may also
be used to further decontaminate the exterior. Do not use abrasive
cleansers or scouring pads, as they may mar the surface. Do not
allow liquid to enter the cooling chamber or other portions of the
instrument.
3. Cleaning the Solenoid Impactor
Freezing the sample depends on the solenoid impactor striking the
sample tip when the preset impact point of 3200 is shown on the
display. Depending on the instrument test usage, the solenoid
impactor should be periodically cleaned to remove any sample
residue and chamber cleaner fibers. Instruments used daily should
be cleaned monthly, while instruments that are used only occasion-
ally should be cleaned every three months. See the solenoid clean-
ing instructions, later in this section.
Fuse replacement
If you determine that your instrument is not functioning because of
blown fuses, you will need to replace the fuses using the following pro-
cedure.
1. Switch the power switch to the off position and disconnect the
power cord.
2. Use a small flat-bladed screwdriver or similar tool to remove the
fuse holder door. Remove the fuse holder.
3. Double-check the values marked on the fuses. The Model 3320 will
automatically adjust for voltages between 100VAC and 240VAC,
32
Содержание 3320
Страница 1: ...The Advanced Micro Osmometer Model 3320 User s Guide 3325 Rev12 041417 ...
Страница 6: ...vi Advanced Micro Osmometer Model 3320 User s Guide Notes ...
Страница 8: ...viii Advanced Micro Osmometer Model 3320 User s Guide Notes ...
Страница 26: ...6 Advanced Micro Osmometer Model 3320 User s Guide Notes ...
Страница 42: ...Advanced Micro Osmometer Model 3320 User s Guide 22 Notes ...
Страница 74: ...Advanced Micro Osmometer Model 3320 User s Guide 54 Notes ...
Страница 80: ...Advanced Micro Osmometer Model 3320 User s Guide 60 Notes ...
Страница 90: ...Advanced Micro Osmometer Model 3320 User s Guide 70 Date Problem Symptom Action ...