1316-900 Revision AA
10kW Precision Calorimeter
25
Table 3: General Troubleshooting Guide
Symptom
Corrective Action
No power displayed
Check load DC resistance for possible failure
Power won’t settle or excessively long settling
time
Check RF amp for possible power drifting
Make sure chiller set point and PID parameters
are correct
Flow rate much less than nominal
Check for blockage in fluid filter
AC calibrated power equals input power
Check
calibration.ini
file for proper values
RF calibrated power equals AC calibrated power
Check
reflection.ini
file for proper values
Extreme ambient temperature readings
Check for fan failure or inlet air blockage
No communication with 1316 Calorimeter
Check for proper IP address