1316-900 Revision AA
10kW Precision Calorimeter
18
Figure 10:
Step 1
group fields
Field
Description
ID
Technician ID or initials.
Procedure
The calibration authority/procedure for
the test.
Report
Number
The report number must be 10
characters and should be unique. The
date, including minutes
(yymmddhhmm), is recommended to
ensure each report number is unique.
Location
Company or department name, or test
station number.
Ambient
Temp
Ambient temperature of the calibration
environment.
Humidity
Ambient humidity in the calibration
environment.
Sensor
Loc
Physical location of the
temperature/humidity sensor
Table 1: Step 1 group field definitions
14.
In the
Calorimeter Check
group, verify the Calorimeter
Flow:
and
Flow Meter:
fields
indicate 4.05 GPM ±0.03, and that
Coolant Temp:
15.
(Optional) In the
Step 2
group, click to select the
Check power to standard
check box
, Step 2).This setting instructs the HPC-CAL calibration software to verify
the connection between the Power Analyzer and AC Power Source.
16.
In the
Step 2
group, click the
Check All Equipment
button (
, Step 3).
17.
Wait approximately 60 seconds for the HPC-CAL calibration software to verify
communication with the connected instruments. Verify the
Power Source
,
Power
Standard
, and
Calorimeter
checkboxes are checked. The
Check All Equipment
button text should be black (
, Step 4).
18.
In the
Step 3
group, click the
Start Auto Test
button (
, Step 5).