1316-900 Revision AA
10kW Precision Calorimeter
16
8.
Connect cables CA-70-5 and CA-71-5 to the Voltage/Current RF Adapter as shown in
Figure 6: CA-70-5 and CA-71-5 connections to Voltage/Current RF Adapter
9.
Connect the Voltage/Current RF Adapter output HN RF connector to the Calorimeter
input HN RF connector.
10.
Allow the system to stabilize for no less than two (2) hours before performing the
remaining steps of this procedure.
Calibration
1.
Launch the HPC-CAL calibration software on the system workstation.
2.
Click the
Test Definition
3.
Click the
Read Test Definition
Figure 7: HPC-CAL calibration software main window
4.
In the
Open File
dialog box, select the Test Definition file titled
HPC-
Cal_TestDefinitions_SNxxxx.XML
.
5.
Click
Open
to load the selected calibration test parameters into the HPC-CAL