ADOS S.R.L. Buccinasco (MI)
492-MTU10006
Rev. 4
Page 29 of 49
4.3.3
Load cell connection
The load cell must be connected to the instrument with a shielded cable according to the attached table. To
this regard, we recommended the use of our shielded cable (Type CS-4).
The cable shield at A100E side must be grounded.
The connection must be made in accordance with the color code reported on the load cell (when using
ADOS cells refer to table 167-MTSCACOC).
The A100E includes the standard connection of load cells with 4 wires. For a connection to load cells
equipped with senses (therefore with a 6-wire connection), contact ADOS.
THE CONNECTION CABLE BETWEEN THE LOAD CELLS AND THE INSTRUMENT
MUST FOLLOW A COMPLETELY SEPARATE LAYOUT, TO AVOID UNWANTED
COUPLING THAT CAN REDUCE THE QUALITY OF MEASUREMENT.
TERMINAL
PIN
FUNCTION
ADOS COLOR CODE
M4
1
+ SIGNAL
GREEN
M4
2
- SIGNAL
WHITE
M4
3
+ POWER SUPPLY
RED
M4
4
- POWER SUPPLY
BLACK
Cable shield must be connected to a good ground reference plane.