Seite
34
von
112
M118000-01-210305-Rev002-UM-en
9. Technical Data
Power supply
Mains operation:
Use only with mains adapter UES06WOCP-060100SPA
Mains voltage:
100 – 240 V AC, 0.2A
Frequency:
50/60 Hz
Supply Voltage:
6 V DC
Current:
1.0 A (max.)
Battery operation:
4 x 1.5 V AA alkaline batteries
(> 5000 measurements / >100 h operating time)
"Weight“ Measuring
range
Single Range
Dual Range
Max. load capacity:
15 kg
20 kg
Division:
5 g
5 g < 10 kg > 10g
Accuracy:
0 ≤ 2,500 kg:
±5g
0 ≤ 2,500 kg:
±5g
2,505 ≤ 10,000 kg:
±10g
2,505 ≤ 10,000 kg:
±10g
10,010 ≤ 15,000 kg:
±15g
10,010 ≤ 20,000 kg:
±20g
"Length“ Measuring range
Measuring Range:
39,8 – 80,2 cm
Division:
0,2 cm
Accuracy:
±0,4 cm
Ambient conditions
Operating temperature:
+10°C to +40°C
Storage and Transport
temperature:
-20°C to +60°C
Humidity:
10% - 95% RH, non-condensing
Air pressure:
700 hPa - 1060 hPa
Device classification
The combination of scale and adapter are specified as ME System. Adapter
is considered as non-ME equipment part of ME System.
Mode of operation:
Continuous operation
Degree of protection:
Type BF applied part (scale top surface)
Battery mode:
Internally powered ME equipment
AC adapter mode:
Class II ME equipment
Software
The software version is displayed when the scale is switched on.
Wireless:
Frequency:
2402 MHz to 2483.5 MHz
Transmission power:
+4 dbm
Range:
10 metres
Housing
Dimensions:
610 x 390 x 140 mm
655-1055 x 390 x 140 mm
without length measuring system
with length measuring system
Net weight:
3,5 kg / 3,7 kg
without / with length measuring system
Material:
ABS plastic
Protection class:
IP21; the device is protected against solid foreign objects with a diameter of
≥
12.5 mm. Protected against vertically falling drops of water or
condensation.
Service life:
The design provides you with a service lifetime of 8 years.