1. Intended Use
Your ADE electronic baby scale is a quality product and manufactured for determining weight (up
to 15 respectively 20 kg) and optionally length (40 to 80 cm) of babies and toddlers, who cannot yet
stand still on a personal scale by themselves without assistance.
The scale may be used in all professional health care institutions, for medical, diagnostic and
recovery purposes that require mandatory calibration.
The maximum capacity of the scale is 15 respectively 20 kg. To achieve precise results, please
read the user guide carefully and follow the instructions contained therein. The scale may only be
operated and maintained by trained personnel.
The device may only be used as intended. All applications of the device not mentioned in the
chapter "Intended use" are considered as improper use. It is the user of the device, but not the
manufacturer, who assumes liability for any resulting damage to property or personal injury
resulting from misuse.
The use of accessories other than the original accessories supplied by the manufacturer may void
this warranty.
Warning:
This device may not be modified without the manufacturer's permission.
During use, do not touch the mains adapter/battery and the baby to be weighed at the
same time.
Do not use the device in an oxygen-rich environment.
2. Safety Instructions
2.1
General Safety Instructions
Be sure to read, understand and follow all instructions in this manual and others that accompany the
system and its components, as well as country-specific installation standards, applicable safety
regulations and accident prevention rules.
- Handle the scale with care and always keep in mind that it is a precision measuring instrument.
- Make sure that the baby to be weighed is centred on the weighing surface.
- Do not use the scale if one or more security marks are damaged.
- Do not use the scale if the calibration counter reading displayed when the scale is switched on
does not match the number marked on the valid calibration counter mark.
- The scale may only be operated and maintained by trained and authorised skilled personnel.
- Before first use, make sure that the mains voltage and current type stated on the name plate
match the mains voltage and current type at the place of use.
- Never touch baby and mains adapter simultaneous.
- Only the power adapters authorised by the manufacturer ADE may be used. Otherwise, there is
a risk that other electrical devices will be affected.
- Only Bluetooth devices authorised by the manufacturer ADE may be connected. Otherwise,
there is a risk that the specified performance level will be compromised.
- Anyone connecting additional equipment or power supply (other than specified in section 8) to
the equipment is responsible that the system complies with the requirements of the standard
IEC 60601-1.
- The plug/adapter connector isolates the device from the mains supply. Do not place the device
in a position where it is difficult to disconnect the power supply to safely stop the operation of
the device.