The every parameter changing like as following list:
From above lists, the transmitter is unqualified before calibration. However, it meets the accuracy requirements after calibration.
Remark:PV and AO calibrations are typical for HART instruments. However, AO instruction aren't always available for some transmitters.
For this situation, the ADT672 has a special calibration method to meet the calibration requirements.
Supplied pressure
PV value
PV range
AO output
CA value
Maximum error
Before calibration
00.000 kPa
16.000 kPa
00.010 kPa
16.020 kPa
4.001 mA
20.020 mA
00.000 kPa
16.000 kPa
4.024 mA
20.061 mA
0.15%
0.38%
Supplied pressure
PV value
PV range
AO output
CA value
Maximum error
Calibration process
00.000 kPa
16.000 kPa
00.010 kPa
16.020 kPa
4.024 mA
20.061 mA
00.010 kPa
16.020 kPa
4.024 mA
20.061 mA
0.15%
0.38%
Supplied pressure
PV value
PV range
AO output
CA value
Maximum error
After calibration
00.000 kPa
16.000 kPa
00.010 kPa
16.020 kPa
4.000 mA
20.000mA
16.020 kPa
4.001 mA
20.002mA
0.006%
0.012%
00.010 kPa
Process 1: PV calibration Process 2: AO calibration