2.2 Electrical measure function
2.2.1 Switch measure items
2.2.2 Electrical measure
2.2.2.1 Current zeroing
2.2.3 Voltage measure
2.2.3.1 Voltage zeroing
2.2.4 Temperature measure
2.2.5 Pressure switch measure
By pressing , switch the measure items by the sequence of current(mA), voltage(V), environment temperature(℃), switch
, counting-down pressure leak test .
Electrical terminal is: ◎mA ◎COM
If it is in the range of (0~ 0.2%FS), please clear the current value to zero by pressing .
Electricity terminal is: ◎V
◎COM
If it is in the range of (0~ 0.2%FS), please clear the voltage value to zero by pressing .
The range is (-20~80)℃ and resolution is 0.1℃. This function is to measure the internal temperature of
.
Electrical terminal is ◎SW ◎COM
Indication icons: is switching off ; is switching on .
There are five triggering modes as listed below: