22
Please refer the figure 2-2-6-1 for pressure leak test process. The maximum
time is 99 hours 59 minutes and 59 seconds. The default is 10 seconds. It can
be set in the main menu, as below:
(1) Please select the (LEAK) and (SET).
(2) As figure 2-2-6-2, please input the time as format
hour: minute: second
,
confirm it and return the main interface by pressing .
In the main interface, pressure value is switched between 5 digits and 6 digits
by pressing .
Electricity terminal is ◎24V
◎COM.
In the main interface, please turn on the DC24V output by pressing
(icon appear). Press again to close the DC24V (icon
disappears). It is suggested to turn off the DC24V function if it is not used.
【
】
2.2.7 Change display esolution
2.3 DC24V output function
2.3.1 Power on/off
r
DC24V
Figure 2-2-6-1: Count-down leak test
count-down time format
[hour:minute
second]
:
icon
stop
starting pressure P1
start to count-down
count-down to 00:00:00
count-down time format
[hour:minute
second]
:
icon
round
icon
stop
restart
starting pressure P1
finish pressure P2
process pressure P
Hour Minute Second
first digit flashed to adjust
Figure 2-2-6-2: Count-down time setting
24V
24V