63
(7) Wrong sequence number explanation
1000 Accept outflow from buffer.
1001 The instruction is being protected now.
1004 Digital character string have non-permitted characters.
1005 Pressure unit is irregular.
1007 Parameter is wrong.
1016 The data can't meet the zeroing requirements.
1017 The parameter quantities is not enough.
1018 Unsupported instruction.
1019 The format of operation password is wrong.
1020 r/w Symbol is wrong.
1021 The file number is out of range.
1023 The shortening pressure unit is wrong.
1024 This pressure unit can't use.
1025 The series address is out of range 1-112.
1026 Baud rate is wrong.
1027 The time parameter of 24V open/close is wrong.
1029 Parameter is too long.
1030 Not contact with HART yet.
(8) HART function's return format
Function 0:
Linked the POLLING successfully
Address: F: MVAL: DEVICE: Manufacture: Equipment type: Equipment ID
Example: 001: F: MVAL: DEVICE: Endress & Hauser: Cerabar M: W ["Q