According to the Figure 8 or 9 and the type of reference sensor, to measure the TC or RTD. Then according to
Figure 6 to measure the switch signal.
2. Create the task
Refer to paragraph 6.2.4.
3. Task setting
Refer to the task setting in paragraph 6.1.1.1.
4. Start the task
For part of the content, refer to Start the task in paragraph 6.1.1.4.
When the DUT moves (open/close) during the task, the calibrator will automatically record the temperature
value at the moment of the action, without manually input.
5. Finish the task
Refer to Finish the task in paragraph 6.1.1.5.
6.3 Calibrate flow devices
6.3.1 Delta- pressure flowmeter
1. Connect DUT
Connect the external pressure module to the module Lemo type connection A or B.
Connect DUT output with the electrical module of the calibrator according to the signal type of DUT output
(current/ voltage), as shown in the Figure 6 or 7.
2. Create the task
Create the corresponding DUT tasks as described above, and fulfill the necessary information. See Table 30.