59
For part of the content, refer to Start the task in paragraph 6.1.1.4.
The calibrator will automatically use the measured value of the voltage measurement, current measurement or the
HART measurement according to the DUT output type set by the user, without manually input.
5. Finish the task
Refer to Finish the task in paragraph 6.1.1.5.
6.1.3 Pressure switch
1. Connect to DUT
Connect the external pressure module to the Lemo type connection A or B. Connect the DUT output with calibrator’s
electrical module. (Figure 6)
2. Create the task
Create the corresponding DUT tasks as described above, and fulfill the necessary information
Table 24 Pressure switch task setting
Subject
Valid value
Description
Name
Numbers, letters, symbols
Task name
S/N
Numbers, letters, symbols
Task S/N
Model
Numbers, letters, symbols
Model of DUT
Pressure type
GP/AP/DP
Pressure type of DUT
Range
Numbers
Input range of DUT
Accuracy
0.5%,1%,1.5%,2%,
2.5%,4%,custom
Accuracy of DUT. If choose custom, the input number means the
accuracy level of this pressure switch.