80
6.4.2 Loop signal source
1. Connect to DUT
Connect the DUT with the measurement module of the calibrator, as shown in the figure 7 (for current
measurement) or 6 (for voltage measurement).
2. Create the task
Create the corresponding DUT tasks as described above, and fulfill the necessary information. See Table 35.
Table 34 Loop signal source task setting
Subject
Valid value
Description
Name
Numbers, letters, symbols
Task name
S/N
Numbers, letters, symbols
Task S/N
Model
Numbers, letters, symbols
Model of DUT
Range
(4~20)mA,(0~10)mA,(0~20)mA,(1~5)V,(0~5)V,(0~10)V, custom
Output range of DUT
Accuracy
0.025%,0.05%,0.1%,0.16%,0.25%,0.4%,1%,1.6%,2.5%,4%,custom
Accuracy of DUT
Transfer
function
Linear / square / square root
Transfer function of the DUT
Owner
Numbers, letters, symbols
Owner of DUT
Location
Numbers, letters, symbols
Location of DUT
Note
Numbers, letters, symbols
Additional information
3. Task setting
Refer to the task setting in paragraph 6.1.1.