56
Table 21 Task running setting
Subject
Valid value
Description
Set point list
The effective value of the set point
pressure depends on the pressure
range of the DUT; up to 17 points
can be set
Set the calibration point for this task, and the calibrator will
automatically set the default calibration point into equal difference
according to the range of the DUT and the number of set points.
However, you can click the "Set Point List" item to manually set them.
At this time, the unit of the set point is consistent with the unit of the
DUT. Use the arrow button on the bottom of the screen, or click the
number of set points and then enter the number of set points through
the keyboard to change the number of set points, or modify a set point
individually.
Cycle count
1,2,3
Cycles during calibration
Strokes
One way/ Round trip
Stroke during calibration
Tap
On/ off
Tap on or off
4. Start the task
Click Start to start the task.
Click button
or
to switch between lines mode and table mode.
Calibration process:
(1) After the pressure is controlled to the first set point by the device, read and enter the displayed value of the
DUT on the task interface.
(2) click Right icon to go to the next point.
Содержание 226
Страница 1: ...226 227 Multifunction Process Calibrator...
Страница 2: ......
Страница 3: ...ADT226 227 Multifunction Process Calibrator User Manual Version 2306V01 Additel Corporation...
Страница 4: ......
Страница 6: ......
Страница 31: ...19 1 4 Basic Structure Figure 1 Basic Structure...
Страница 33: ...21 Figure 2 Power adaptor...
Страница 64: ...52 External module information including external module A and external module B...
Страница 107: ...95...