35
RF emissions CISPR 11
Group 1
Emissions
Test
Electromagnetic Environment -
Guidance
Guidelines and manufacturer's declaration - electromagnetic emissions
The e-sphyg 3 Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of e-sphyg 3 Blood Pressure Monitor should assure that it is used in such an environment.
Compliance
The e-sphyg 3 Digital Blood Pressure Monitor uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
The e-sphyg 3 Digital Blood Pressure Monitor is suitable for use in all
establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
1
2
3
4
5
6
7
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV
differential mode
± 2 kV common
mode
Immunity
Test
IEC 60601
Test Level
Electromagnetic
Environment - Guidance
Guidance and manufacturer’s declaration – electromagnetic immunity
The e-sphyg 3 Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the e-sphyg 3 Digital Blood Pressure Monitor should assure that it is used in such an environment.
Note - UT is the a. c. mains voltage prior to application of the test level.
Compliance
Level
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrostatic
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
3 A/m
3 A/m
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
< 5 % UT
(>95 % dip in UT )
for 0.5 cycle
40 % UT
(60 % dip in UT )
for 5 cycles
70 % UT
(30 % dip in UT )
for 25 cycles
< 5 % UT
(>95 % dip in UT )
for 5 sec
< 5 % UT
(>95 % dip in UT )
for 0.5 cycle
40 % UT
(60 % dip in UT )
for 5 cycles
70 % UT
(30 % dip in UT )
for 25 cycles
< 5 % UT
(>95 % dip in UT )
for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
e-sphyg 3 Digital Blood Pressure Monitor requires
continued operation during power mains
interruptions, it is recommended that the e-sphyg 3
Digital Blood Pressure Monitor be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Annex of Report
Manufacturer’s Declaration of the EUT
Report No.: TRE14120020
A 2
Issued: 2014-12-15
Guidance and manufacturer’s declaration – electromagnetic emission – for all EQUIPMENT AND SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS