Each cuff is provided with 50 cm air tube.
Use only cuffs provided by ADC!
Contact ADC or an authorized distributor to purchase cuffs.
Measurement Procedure
Patient position
The patient should sit or lie comfortably. The arm should be fully supported on
a flat surface at heart level. (If the arm’s position varies, or is not level with the
heart, measurement values obtained will not be consistent with the patient’s
true blood pressure.) When seated, the patient should have their back and arm
supported, and their legs should not be crossed. Prior to measurement the
patient should relax comfortably for five minutes and should refrain from
talking or moving during measurement.
Apply the cuff
Place the cuff over the bare upper arm with the
artery mark positioned directly over the brachial
artery. The bottom edge of the cuff should be
approximately one inch (2-3 cm) above the
antecubital fold. Wrap the opposite end of the cuff around the arm snugly
and smoothly and engage adhesive strip.
To verify a correct fit, check that the Index Line falls between the two Range
Lines.
Left Arm
Tube