43
Default mode is 2WIN-S (Refraction) mode. When KALEIDOS App is restarted 2WIN (Refraction) mode
is loaded.
CR Exam Procedure
CR exam consists of a sequence of three acquisitions: one binocular and two cover tests under an infrared
occluder. The patient must look at the white LED that turns on during the measurement.
To perform a Corneal Reflexes measurement, follow these steps:
1.
Select CR mode, as described above.
2.
Tap on
“
Start a new measurement
”
.
3.
Enter the patient
’
s name and date of birth. If needed, patient data insertion can be skipped.
4.
Click on
“
Start
”
and the Corneal Reflexes measurement window opens.