60
Appendix B
–
EMR input and output files
specifications
This appendix describes how to implement the EMR interface in order to exchange patient data and
measurement/screening results between your EMR system and your 2WIN-S device.
The EMR interface needs to be implemented by your EMR System provider. Please note that neither Adaptica
nor any Adaptica distributor can perform the needed modifications in your EMR system.
This is a
“
full
”
integration, i.e., patient name or ID is sent from the EMR system to 2WIN-S, and
measurement/screening results are saved to data fields within the patient
’
s EMR.
Definitions
•
Shared Folder:
The Shared Folder can be used to exchange input and output files between your EMR
system and your KALEIDOS App.
This folder must be created in the EMR system PC and must be shared on the LAN (Local Area
Network). The folder must be accessible by your Control Android Tablet, which must be connected
to the same network, and by your KALEIDOS App.
Use different Shared Folders for different 2WIN-S.
•
Input file
: A file generated by your EMR system to provide patient information prior to a
measureme
nt/screening. This file must be named “
emr_input.csv
”, and it
must be saved in the
Shared Folder.
•
Output file
: A file generated by the KALEIDOS App after a measurement/screening that has been
performed or has been aborted (i.e., an inconclusive measurement/ screening). This file is named
“
emr_output.csv
”, and it can be retrieved in the Shared Folder.
Input file specifications
•
Patient information must be put in a CSV File, side by side in one single line and without a space after
the separator.
•
The sequence for patient information must be the following: last name, first name, date of birth
(YYYY-MM-DD), patient ID, location, contact, autostart.
•
Autostart field can be 0 or 1. If autostart=0, when the input file is created, KALEIDOS App opens
patient data window. The user must press START to start the measurement. If autostart=1, when the
input file is created, the measurement automatically starts, without pressing the START button.
Example:
Smith,John,2016-02-24,1234567890,Downtown Clinic,johnsm[email protected],0
is a valid input file.
Содержание 2WIN-S
Страница 1: ...1 2WIN S 2WIN S Plus OPERATOR MANUAL ...
Страница 14: ...14 BACK INDOOR CONTROL PANEL UPPER SIDE ...