60
|
P a g e
© Adam Equipment Company 2017
11.5 ANALYZER
CONFIGURATION
THROUGH
THE
SERIAL
INTERFACE
The analyzer can use the serial interface to set parameters for weighing test
and for analyzer configuration. The parameters are set by sending a line of
data that has the description and the setting desired.
It is possible to create a text file that contains all the data necessary to set he
test conditions for a particular test (or analyzer configuration) and send this to
the analyzer. The commands are not case sensitive, i.e. parity=0 and
PARITY=0 are both accepted.
11.5.1
TEST PARAMETERS
The parameters that control a test are listed below. To change a test
parameter it is necessary to send a command for the parameter to modify
along with the value to change.
For example to change the test parameters to use a single heat and set the
temperature to 115ºC send 2 commands.
Heat=0<cr><lf>
Single Temp=115<cr><lf>
All commands are in English regardless of the language selected for the
analyzer operation. The commands can be sent in both upper case letters
and lower case letter, i.e. “!KT” or “!kt” are both acceptable.
Moisture Result=1;
#0=%M, 1=%S, 2=%Atro_M, 3=%Atro_S default=0;
Heat=1;
#0=Single, 1=Ramp, 2=Step
default=0;
Single
Temp=100;
#050C
to
160C,
default=100;
Step temp1=050;
#050C to 160C,
default=100;
Stem Time1=30:00;
#Time 03:00 to 99:00
default=30:00;
Step temp2=080;
#050C to 160C, 000C is off
default=100;
Stem Time2=30:00;
#Time 03:00 to 99:00
default=30:00;
Step temp3=050;
#050C to 160C, 000C is off
default=100;
Stem Time3=30:00;
#Time 03:00 to 99:00
default=30:00;
Ramp
Temp=100;
#050C
to
160C,
default=100;
Ramp Time=30:00;
#Time 03:00 to 99:00
default=30:00;
Rapid=0;
#Rapid heat 0=Off, 1=On, 2=Manual
default=0;
Interval Time=20;
#05 sec to 99 sec
default=15;
Stop mode=0;
#0=Time, 1=Stable, 2=Stab+Time
default=3;
Stop time=10;
#time = 03 to 99 min
default=20;
Stable Range=0.010;
#grams default=0.010 for PMB 53, 0.10 for PMB 202;
Stable
Time=10;
#seconds,
01-99
default=02;
Start
mode=0;
#0=Manual
1=Auto
default=0;
savepos=01;
#memory location 01 to 49;