34
|
P a g e
© Adam Equipment Company 2017
If the rapid heat function has been set to MANUAL the display will ask if the
user wants the rapid heat function to operate during this test.
The display will show the last setting used, either:
Use the
[Dspl/
]
or
[Test/
]
keys to change the option, then press
[Enter]
to continue the test.
The top line will show the time elapsed time of the test and the chamber
temperature. The main display will show the current result, updated at the
interval time selected when setting test parameters.
If the Rapid heat is active a “RA” will be displayed in front of the temperature
reading.
An “A” will be on the left side of the display if %Moisture /Dry or %Solid/Dry is
the result selected. The %M or %S will be on the right side of the display
showing %Moisture or %Solid.
The display will continue to be updated at the time interval set for the test.
During the test the
[Dspl/
]
key can be pressed to see other information
such as current mass, other results such as %Solid, time remaining if time
limit is set for the test. Each alternate display is shown for 5 seconds then
return to normal display, or if
[Dspl/
]
is pressed again go to the next
alternate display.
OFF
rapid
ON
rapid
03:45 RA 100
ºC
16.12
%