The balance has a number of technical parameters the user can adjust.
Display Description
Default Value
n FIL
Filter settings - 1 to 3
1 will allow the display to update at it’s fastest
1 FIL
n ZEO
The Automatic Re zero range - 1 to 7
1 is the minimum A/Z range and 7 is the maximum
4 ZEO
n STA
Stability symbol range - 1 to 6
1 is the smallest stability range and 6 is the largest.
2 STA
1) Switch on the balance then press and hold the
[Unit/Clr]
key during the self-
checking test of the display.
2) After a few seconds, the display will show the first function
F1 UNT
.
3) Keep pressing the
[Air/Liquid]
key until
F8 SET
is displayed.
4) Press the
[Tare]
key to view the current setting.
5) To change the settings press the
[Air/Liquid]
key. Press the
[Tare]
key to store
the setting and advance to the next parameter. When complete the display will
show the
TECH
parameter.
The display will now show
TECH
. This parameter is for qualified technicians only and
requires entry via a pass code. Press the
[Unit/Clr]
key to return to weighing or
press
[Air/Liquid]
to go to the first function again.