7.7
ADJUST
THE
VALUE
OF
THE
INTERNAL
MASS,
F7
F7
F7
F7
C
C
C
CA
A
A
A
The internal mass value stored in memory can be adjusted if required.
1)
Switch on the balance then press and hold the
[Unit/Clr]
key during the self-
checking test of the display.
2)
After a few seconds, the display will show the first function
F1 UNT
.
3)
Press the
[Air/Liquid]
key until
F7 CA
is displayed.
4)
Press the
[Tare]
key to view the current setting.
5)
The display will show the current value with the first digit
flashing. To change the flashing digit position press the
[Hold/Print] key.
Press the
[Air/Liquid] key
to increase the value
or the
[Density] key
to decrease the value. The internal calibration weight
should only change between
95.00
and
105.00
grams or
495.00
and
505.00
grams (depending upon the model).
6)
Press
[Tare]
to confirm and move to the next setting.
7)
When
F8 SEt
is displayed press the
[Tare]
key to enter the parameter, the
[Unit/Clr]
key to return to weighing, or press
[Air/Liquid]
to go to the next
function.
*Changing the value of the internal mass will affect the weighing readings following
an internal calibration. To check if your internal mass value is correct you should
only use high quality weights that are accurate to the readability of the balance.