20
© Adam Equipment Company 2012
7.3
BACKLIGHT,
F3
F3
F3
F3
b
b
b
bLLLL
The backlight may be enabled or disabled by the user. If the backlight is disabled the
battery life will be greater. The following settings are available:
AU bL
Sets the backlight to operate automatically when a weight is
placed on the balance or a key is pressed.
On bL
Sets the backlight to be on at all times.
Off bL
Sets the backlight to be off at all times.
1) Switch on the balance then press and hold the
[Unit/Clr]
key during the self-
checking test of the display.
2) After a few seconds, the display will show the first function
F1 UNT
.
3) Press the
[Air/Liquid]
key to select
F3 bL
.
3)
Press the
[Tare]
key to view the current settings for the backlight.
4)
To change the setting press the
[Air/Liquid]
key
to
scroll through and select a
new setting as shown above.
5)
Press
[Tare]
to confirm and move to the next setting.
6)
When
F4 Ser
is displayed press the
[Tare]
key to enter the parameter, the
[Unit/Clr]
key to return to weighing, or press
[Air/Liquid]
to go to the next
function.
7.4
PRINTING
PARAMETERS,
F4
F4
F4
F4
S
S
S
Se
e
e
errrr
1)
Switch on the balance then press and hold the
[Unit/Clr]
key during the self-
checking test of the display.
2)
After a few seconds, the display will show the first function
F1 UNT
.
3)
Keep pressing the
[Air/Liquid]
key until
F4 SEr
is displayed.
4)
Press the
[Tare]
key to view the current setting.