14
© Adam Equipment Company 2012
6.0
DENSITY
PARAMETERS
To use the density function it is necessary to set parameters describing the liquid
used and the alloy being tested. These parameters can be set when the balance is in
the normal weighing mode.
During normal weighing with the display showing zero weight, press the
[Density]
key. If the display does not change to show
Liquid
check the parameter for the Hold
Function (section 7.2). If it has been set to off “
Hd Off”,
density is disabled and this
section will not operate. Set to Hold On, “
Hd On”,
or Hold Automatic, “
Hd AU”,
to
enable the density function.
The display will show
Liquid
to identify the section for setting parameters associated
with the liquid used.
Press
[Tare]
to enter this section or press
[Air/Liquid]
to view the
Alloy
parameters.
When Alloy is displayed press
[Tare]
to enter this section or press
[Air/Liquid]
to
return to the
Liquid
parameters.
6.1
LIQUID
AND
TEMPERATURE
PARAMETERS
The parameters available are water (H2O) and any other user definable liquid. If
water is used and selected the next parameter will ask to determine the
temperature of the water, selections are either using the automatic temperature
sensor or to manually input the temperature.
If another liquid is used then the user must set the density of that liquid.
The display will show
Liquid
, Press
[Tare]
to enter this section.
Press
[Air/Liquid]
or
[Density]
to select
H2O
or
Define den
. Press
[Tare]
to select the
setting or press
[Unit/Clr]
to return to the
Liquid
display.
If
H2O
is selected the display will show the current setting for the temperature
determination, either
Auto tmp
, or
Set tmp
.
Press
[Air/Liquid]
or
[Density]
to select the option, then press
[Tare]
to store the
selection.