EN
© Adam Equipment Company 2018
19
3) Keep pressing the
[Mode]
key until
F4 oFF
is displayed.
4) Press the
[‐>0/T<‐]
key to view the current settings.
5) Press
[Mode]
to change the settings (0, 5, 10, 20 and 30 minutes).
6) Press
[‐>0/T<‐]
to store a particular setting. The display will return to
F4 oFF
7) When
F4 oFF
is displayed press the
[Print]
key to return to weighing or press
[Mode]
to go to the next function.
6.5
SELECTING
THE
INTERNAL
OR
EXTERNAL
CALIBRATION
(
N
OT POSSIBLE ON
APPROVED
M
ODELS
)
You can select if the internal mass or an external mass is to be used to calibrate the
balance.
1) To set this parameter press the
[Mode]
key during self‐checking test when
turning on the balance.
2) After a few seconds, the display will show the first function
F1 UNT
.
3) Keep pressing the
[Mode]
key until
F5 IEC
is displayed.
4) Press the
[‐>0/T<‐]
key to view the current settings.
5) To change the settings press the
[Mode]
key
to change from
INt
(internal
calibration) or
E
(external calibration)
6) Press
[‐>0/T<‐]
to store a particular setting. The display will return to
F5 IEC
7) When
F5 IEC
is displayed press the
[Print]
key to return to weighing or press
[Mode]
to go to the next function.