7.0 CALIBRATION
(
NOT AVAILABLE ON APPROVED VERSIONS
)
The
Highland
series of balances comes standard with
HandiCal
TM
internal calibration
to make calibrating the balance quick and easy. However you can also calibrate the
balance using an external verification weight if needed. The
HandiCal
method is the
default but if you would like to use external calibration then you must first enable
this via the parameter (section
6.5 Selecting The Internal Or External Calibration
).
Install the handle for the calibration as shown below. To calibrate the handle is
pushed to the side to lower the calibration mass.
INTERNAL CALIBRATION USING HANDICAL
TM
1)
Press the
[on/off]
key to turn the power on.
2)
Press the
[Smpl]
and
[Print]
keys at the same time during the self‐checking
test.
3)
The display will show
unload.
Remove any weight from the stainless steel pan.
4)
When the stable indicator shows press the
[‐>0/T<‐]
key.
5)
The display shows
C Int.
Press the
[‐>0/T<‐]
key
6)
The display shows
LoAd.
Lower the internal calibration mass located behind
the pan. Use the handle to help lower the weight as far as it will go. Press the
[‐>0/T<‐]
key
once the stable sign is shown.
7)
The display will show
PASS.
Rotate the weight back to its off position. Once
removed the balance will return to zero.