34 |
P a g e
© Adam Equipment Company 2017
6.0 SECURE
PARAMETERS
6.1
SETTING
UP
SECURED
PARAMETERS
The following parameters can only be accessed when the service access
code 9999 is used.
Turn the scale on while holding the
[Total]
key down and then release the
key.
Key in the Service Access Code as described in section 5.1.1.
The first function shown is
F1
, the Baud Rate Setting.
“
F1
”
“
bAud
” “
SEt
”
This procedure gives access to the following parameters:
F1
Baud Rate Setting
F2
Parity setting
F3
Sleep
F4
Weight Calibration
F5
Access Code Number Setting
F6
Language
F7
Enable the
[€/£]
key function.
F8
Denomination Setting
F9
Linearity Calibration
F10 Factory Calibration Values
See Section 5.4 for details of Functions F1 to F6.
These parameters are
set as in section 5.4.
These parameters are
only accessible when
the service access code
is used.