Place the calibration weight on the scale.
Press the
[Total]
key to show the ADC values for this weight.
“
LoAd
” “
10
”
“
345689
”
Press the
[Total]
key to show the ADC values for this weight. The
display will then show:
“
SPAN
” “
PASS
” “
”
After few seconds it will show the next function.
“
F5
” “
ACCES
” “
CAL
”
Remove the weight from the scale.
If
[
Z/T
]
Esc
is pressed during calibration return to normal operation without
doing the calibration.
5.4.5
F5
‐ Access Code Number
To change the user access code number used to enter the section of the
program for changing coin weights (see Section 5.2) follow this procedure.
This parameter can only be accessed by skipping the F3 Calibration function
by using the
[MC]
key.
When the display is showing:
“
F5
” “
ACCES
” “
SEt
”
Press the
[Total]
key. The display will show: