7.0 ERROR
MESSAGES
When power is first applied to the scale the software does a self check and
then verifies the voltages from the load cell are acceptable. The display will
show error messages if the voltage is not within certain limits. If these error
messages appear contact your supplier for assistance.
ERROR
CODE
DESCRIPTION
POSSIBLE CAUSES
Err 1
Error setting a date
Trying to set a date that is not
realistic.
Err 2
Error setting time
Trying to set a time that is not
realistic.
Err 4
Initial Zero is greater than
allowed (typically 4% of the
maximum capacity) when
power is turned on or when
the
[
Z/T
]
Esc
key is
pressed,
Weight is on the pan when
turning the scale on.
Excessive weight on the pan
when zeroing the scale.
Improper calibration of the
scale.
Damaged load cell.
Damaged Electronics.
Err tYPE
Tokens are not allowed to be
added to memory.
Shown when trying to add
tokens to accumulation
memory.
SPAn FAIL HI
SPAn FAIL Lo
Calibration Error
Will show Hi or Lo if the
calibration mass is too large or
small to allow correct
calibration.
FAIL
Password failed
Password was not correct.