Acromag IntelliPack 841T Скачать руководство пользователя страница 10

IntelliPack Series 841T Transmitter/Alarm User's Manual                                                                Frequency Input 
___________________________________________________________________________________________ 

 - 10 - 

 
Note: 

 For sensitive applications, high frequency noise may be 

reduced by placing a 0.1uF capacitor directly across the load. 

 

Voltage Output Specifications: 

Output Range: 

 0-10V DC, 0-5V DC. 

Output Accuracy:

  See Table 7. 

Output Current:

  0-10mA DC maximum. 

Output Impedance:

  1

Output Resolution:

  See Table 7. 

Output Short Circuit Protection:

  Included 

Current Output Specifications: 

Output Ranges:  

0-20mA DC, 4-20mA DC, or 0-1mA DC. 

Output Maximum Current:

  21.6mA typical. 

Output Accuracy:

  See Table 7. 

Output Compliance:

 10V Minimum, 12V Typical. 

Output Resolution:

  See Table 7. 

Output Load Resistance Range:

  0 to 550

, typical. 

 
Table 7: Analog Output Range Resolution & Accuracy 

 
Output Range

3

 

 
Resolution 

Accuracy

1,2  

at 25

C

 

Percent-of-Span 

 

 

Output 

Overall

4

 

0 to 20mA DC 

0.0025% 

0.025% 

0.050% 

4 to 20mA DC 

0.0025% 

0.025% 

0.050% 

0 to 1mA DC 

0.0370% 

0.100% 

0.125% 

0 to 10V DC 

0.0025% 

0.025% 

0.050% 

0 to 5V DC 

0.0040% 

0.050% 

0.075% 

 
Notes (Table 7): 

1.  

Voltage outputs unloaded.  Loading will add “I*R” error. 

2.   Software calibration produces high accuracy. 
3.   All current ranges are subsets of a 0-24mA DAC range 

which allows for under and over range capability.  The 0-
5V range is a subset of the 0-10V range. 

4.   For overall accuracy (input to output) add 0.025% to 

output accuracy, assuming full input range is utilized. 

 
Digital-to-Analog Converter:

  Analog Devices AD420AR-32, 16-

bit 

-

.  Monotonic to 16 bits. 

DAC Integral Non-Linearity:

 

0.002% (

1.4LSB) of span typical, 

0.012% (

7.9LSB) of span maximum, utilizing full DAC output 

span (0-24mA). 

DAC Output Temperature Drift:

  Better than 

20ppm/

Typical, 

50ppm/

C Maximum. 

Output Update Rate:  

Frequency inputs:

 

1 to 10 per second 

(varies with input frequency).  Pulse counting inputs: 10 per 
second. 

Output Settling Time:

 10ms to 0.1% of full-scale, varies with 

load. 

 
 

RELAY OUTPUT SPECIFICATIONS 
 

Model 841T-1500 units include a SPDT electromechanical 

alarm relay which provides a set of high reliability, Form C 
(Normally Open and Normally Closed), SPDT contacts.  The 
relay operates as a limit alarm with control of deadband/ 
hysteresis, latching/non-latching contacts, failsafe/non-failsafe 
modes, plus relay time delay. 

 

SPDT Alarm Relay Specifications (841T-1500 Units Only): 

Electrical Life - CSA Ratings:

 

25VDC, 5A, 10

5

 operations, resistive. 

48VDC, 0.8A, 10

5

 operations, resistive. 

240VDC, 0.1A, 10

5

 operations, resistive. 

120VAC, 5A, 3x10

4

 operations, resistive. 

240VAC, 5A, 3x10

4

 operations, resistive. 

Note:

 To control a higher amperage device, such as a pump, 

an interposing relay may be used (see Interposing Relay 
Connections Drawing 4501-646). 

Contact Material:

  Silver-cadmium oxide (AgCdO). 

Initial Dielectric Strength:

  Between open contacts: 

1000VAC rms. 

Expected Mechanical Life:

  20 million operations.  External 

contact protection is required for use with inductive loads 
(see Contact Protection Drawing 4501-646). 

Relay Response:

  For an input step change from 10% of 

span on one side of an alarm point to 5% of span on the 
other side of the alarm point, the relay contacts will switch 
tracking the input acquisition time (see Input Acquisition Time 
specification).  If configured with additional Relay Delay Time, 
this will add to the relay response time. 

 
 

ENCLOSURE/PHYSICAL SPECIFICATIONS 

 

See Enclosure Dimensions Drawing 4501-642.  Units are 

packaged in a general purpose plastic enclosure that is DIN rail 
mountable for flexi

ble, high density (approximately 1” wide per 

unit) mounting. 
 

Dimensions

:  Width = 1.05 inches, Height = 4.68 inches, Depth 

= 4.35 inches (see Drawing 4501-642). 

DIN Rail Mounting (-xx0x):

  

DIN rail mount, Type EN50022; “T” 

rail (35mm). 

Connectors:

  Removable plug-in type terminal blocks;  

Current/Voltage Ratings:  15A/300V; Wire Range: AWG #12-
24, stranded or solid copper;  Separate terminal blocks are 
provided for input, power, output, and relay contacts.  For 
supply connections, use No. 14 AWG copper wires rated for 
at least 75

C. 

Case Material:

  Self-extinguishing NYLON type 6.6 polyamide 

thermoplastic, UL94 V-2, color beige; general purpose NEMA 
Type 1 enclosure. 

Printed Circuit Boards:

  Military grade FR-4 epoxy glass. 

Shipping Weight: 

1 pound (0.45 Kg) packed. 

 
 

APPROVALS (-xxx0) 

 
0

:     Agency Approvals - cULus Listed, UL File E199702.   

 

Product approval is limited to general safety requirements of 

the above standards. 

 
Warning:

  This product is not approved for hazardous location  

                  applications. 

 

Содержание IntelliPack 841T

Страница 1: ...Frequency or Pulse Counting Input USER S MANUAL ACROMAG INCORPORATED 30765 South Wixom Road Wixom MI 48393 2417 U S A Tel 248 295 0880 Fax 248 624 9234 Copyright 1997 Acromag Inc Printed in the USA D...

Страница 2: ...T are registered trademarks of Microsoft Corp 1 0 INTRODUCTION These instructions cover hardware functionality of the transmitter models listed in Table 1 Supplementary sheets are attached for units w...

Страница 3: ...lity of this device to be reprogrammed many times Convenient Field Reprogrammability Adjustment to transmitter zero and full scale plus alarm setpoint and deadband can be made via module push buttons...

Страница 4: ...he common DB9P serial port connector of a host computer The adapter also has a 6 wire RJ11 phone jack to connect to the IntelliPack alarm module via a separate interconnecting cable described below Re...

Страница 5: ...ages from the rail Electrical Connections Terminals can accommodate wire from 12 24 AWG stranded or solid copper Strip back wire insulation 1 4 inch on each lead before installing into the terminal bl...

Страница 6: ...tage meter is required to monitor the output level This meter must be accurate beyond the module specifications for best results Note The module s input frequency range must already be set via the Int...

Страница 7: ...d Configuration LED Program Indication LED INDICATOR Constant ON FLASHING Yellow Zero Full Scale labeled Z FS Zero Full Scale 841T 1500 Only Yellow Relay labeled RLY High or Low Setpoint High or Low D...

Страница 8: ...rking and powered serial interface adapter Acromag s Application Engineers can provide further technical assistance if required When needed complete repair services are available from Acromag 6 0 SPEC...

Страница 9: ...tion near the upper end of the range thus decreasing effective accuracy in this range Table 6 Input Measurement Alarm Accuracy Nominal Input Range Accuracy at 25 C Accuracy Over Temperature 0 to 100 0...

Страница 10: ...tacts The relay operates as a limit alarm with control of deadband hysteresis latching non latching contacts failsafe non failsafe modes plus relay time delay SPDT Alarm Relay Specifications 841T 1500...

Страница 11: ...static Discharge ESD Immunity Complies with IEC1000 4 2 Level 3 8KV 4KV air direct discharge to the enclosure port and European Norm EN50082 1 Surge Immunity Complies with IEC1000 4 5 Level 3 2 0KV an...

Страница 12: ...t Range Unit can be configured for either a voltage or current output range A jumper must also be installed between the output I and JMP terminals for voltage output remove this jumper for current out...

Страница 13: ...m state or non failsafe operation relay energized in alarm state Failsafe mode provides the same contact closure for alarm states as for power loss while non failsafe mode uses alarm contact closure o...

Страница 14: ...SIENT SUPPRESSION DEVICES NOT SHOWN BRTN INP UT 2 3 2 2 FQ IN 5 V INPUT SIGNAL CONDITIONING HYSTERESIS DIV PD BIPO PUP COMPARATOR SERIES 841T SIMPLIFIED SCHEMATIC UNIP REF 1 5 V INPUT ISOL ATION MODE...

Страница 15: ...ency Pulse Counting Counts Counts ANALOG INPUT MODEL 811T 1500 ONLY PV PV Percent SEN SOR INPUT BLOCK RANGE ADJUST BLOCK PV Transfer Function Linear Proportional Configuration Variables High or Low Li...

Страница 16: ...IntelliPack Series 841T Transmitter Alarm User s Manual Frequency Input ___________________________________________________________________________________________ 16...

Страница 17: ...0 T O 3 6 VDC REL AY CO NNECTIO NS Acromag GROUND T B4 M O D E RE L AY P WR 8 4 1 T 1 5 0 0 REMO TE EXTERNAL CO UNTER RESET SW I T CH OPEN COL L ECT OR SEE T B2 INPUT CONNECT IONS ON PAGE 2 PRACT ICES...

Страница 18: ...T NOT E 2 T HIS GROUND CONNECT ION IS RECOMMENDED F OR BEST USE CAUT I ON AND AVOID MAKING ADDIT IONAL GROUND CONNECT IONS W HICH COUL D GENERAT E GROUND L OOPS AND MEASUREMENT ERROR T B2 E X C E X C...

Страница 19: ...21 24 CONTACT FIGURE A 46 TB4 TB3 GROUND CONTACT FIGURE B RELAY 46 43 45 N O COM 35 42 JMP TB3 PWR 32 34 31 DC V DC 11 14 12 TB2 22 15 25 23 26 DIODE DC LOAD TB4 N C MOV AC LOAD TB4 N C SPDT CONTACTS...

Страница 20: ...________________________________ 20 RUN Acromag RLY S T Z FS 118 9 4 68 95 3 3 75 MODE S E T D IN EN 50022 35mm T R AIL D IN MOU N TIN G C L 2 34 59 4 SC R EWD R IVER SLOT FOR R EMOVAL FR OM T R AIL 4...

Страница 21: ...______________________________________ 21 Revision History The following table details the revision history for this document Release Date Version EGR DOC Description of Revision 03 AUG 2017 J CAP JAA...

Отзывы: