SmartReader Plus Data Logger
46
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
1.8.4
Specifications
Accuracy:
±0.5% of range plus resolution
Thermocouple
Range
Resolution
Equation
J
-50 to 600°C
(-55 to 1100°F)
0.30°C (0.55°F)
50
K
-100 to 1150°C
(-145 to 2100°F)
0.50°C (0.90°F)
52
T
-200 to 400°C
(-325 to 750°F)
0.40°C (0.75°F)
54
S
0 to 1450°C
(32 to 2642°F)
1.0°C (1.8°F)
74
E
-50 to 680°C
(-58 to 1250°F)
0.24°C (0.44°F)
90
Specifications are subject to change without notice.
1.8.5
Manual Calibration
A simple way to manually adjust the calibration values on a SmartReader Plus 5 thermocouple
logger follows. For each active channel:
1.
Install the thermocouple and set the proper equation number for that thermocouple
type. Set the TrendReader software into Realtime mode.
2.
Using a valid low reference temperature (as low as possible), observe the Realtime
reading and adjust the Low calibration value until the logger displays the reference
temperature.
3.
Using a valid high reference temperature (as high as possible), observe the Realtime
reading and adjust the Mid calibration value until the logger displays the reference
temperature.
To obtain the best results from the calibration, set the reference temperatures as far apart as
possible.
Содержание SmartReader Plus
Страница 3: ...SmartReader Plus is a Registered Trademark of ACR Systems Inc TRADEMARK NOTICES ...
Страница 8: ...SmartReader Plus Data Loggers IV Copyright 2010 ACR Systems Inc All Rights Reserved Index 105 ...
Страница 96: ...SmartReader Plus Data Loggers 87 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Страница 97: ...SmartReader Plus Data Logger 88 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Страница 98: ...SmartReader Plus Data Loggers 89 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Страница 116: ......