SmartReader Plus Data Logger
40
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
and humidity channels. To connect the EH-020A probe to your SmartReader Plus 4, follow
instruction notes packed with each RH probe set -- see Figure 4-2.
Remote Thermistor Temperature Probe
You can monitor temperature remotely using an ET series temperature probe with your
SmartReader Plus 4 by connecting the probe's two wires to the Temperature and Common
terminals. Make sure the External Temperature channel is activated and the proper equation
is selected.
1.6.3
Pressure Calibration
The pressure sensor, although subject to accuracy degradation when exposed to
contaminants and/or extreme environmental conditions, will drift typically less than 1%/year
(under clean conditions). You should check your SmartReader Plus 4 periodically and, if
necessary, recalibrate the pressure sensor. In the event that you wish to confirm the pressure
calibration of your SmartReader Plus 4 logger, you can perform your own calibration using the
adjustment provisions explained in TrendReader software.
The recommended method of calibration is to use a pneumatic pressure calibrator with a
preferred accuracy of at least 0.1% full scale.
Procedure
The procedure to calibrate is as follows:
1.
Activate the pressure and temperature monitoring channels on your SmartReader Plus 4
if not already enabled.
2.
Set the LOW, MID and HIGH calibration values to zero.
3.
With the input port on the pressure sensor left unconnected (open to ambient), record the
pressure reading you see (it should be close to zero if you are using a gauge sensor or
close to atmospheric pressure if you are using an absolute sensor).
4.
Subtract the pressure reading you recorded in step 3, from zero for the gauge sensor or
from calibrator reading for the absolute sensor, and enter it for the Low calibration
adjustment.
5.
Connect the logger's pressure sensor to the calibrator, ensuring all connections are
completely airtight.
6.
Adjust the calibrator to produce a test pressure equal to the full scale range of the logger
and record the pressure reading.
7.
Subtract the pressure reading you recorded in step 6, from the full scale range of the
sensor (the calibrator reading should be the same), and enter it for the Mid calibration
adjustment.
8.
Save the calibration changes to the logger.
NOTE: The value of the Ext. column is factory adjusted and should not be
changed.
Содержание SmartReader Plus
Страница 3: ...SmartReader Plus is a Registered Trademark of ACR Systems Inc TRADEMARK NOTICES ...
Страница 8: ...SmartReader Plus Data Loggers IV Copyright 2010 ACR Systems Inc All Rights Reserved Index 105 ...
Страница 96: ...SmartReader Plus Data Loggers 87 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Страница 97: ...SmartReader Plus Data Logger 88 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Страница 98: ...SmartReader Plus Data Loggers 89 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Страница 116: ......