SmartReader Plus Data Loggers
67
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
1.14
Conservation Logger Temperature, RH and 5V Input Logger
The Conservation Logger is a self-contained temperature and relative humidity logger with two
external channels for use with a variety of sensors.
Figure 11-1: Conservation Logger
1.14.1
Description
The Conservation Logger is designed to monitor temperature and humidity (internal or
external) and determine if the readings measured are out of tolerance based on preset
thresholds. It is equipped with an LED light that will flash its alarm if measurements raise
above or below set limits.
Alarm Flash Mode
When the Alarm Flash mode is selected the logger will check the current sample and
determine if it is above or below the high or low alarm threshold values. If the sample is out of
tolerance then the logger will start a delay timer to allow the environment to return to normal
limits. If the environment does not return to normal limits the alarm will be triggered and the
Alarm LED will begin to flash.
Once an alarm is triggered it is locked and will not clear until the logger is cleared or set up.
Flash Delay
The Flash Delay is the maximum allowable period of time (based on sample periods) for a
channel to be out of tolerance before alarming. The Flash Delay is started when a channels
current sample is found to be out of the preset alarm limits. Once the delay timer is started it
will count down from the preset number of samples to zero. When zero is reached the alarm
will be locked and the Alarm LED will begin to flash.
The Flash Delay timer will be reset if the channel that reported the alarm returns to normal
operating conditions (within the preset tolerances) before the timer reaches zero.
If a threshold is crossed again the delay timer will restart.
Содержание SmartReader Plus
Страница 3: ...SmartReader Plus is a Registered Trademark of ACR Systems Inc TRADEMARK NOTICES ...
Страница 8: ...SmartReader Plus Data Loggers IV Copyright 2010 ACR Systems Inc All Rights Reserved Index 105 ...
Страница 96: ...SmartReader Plus Data Loggers 87 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Страница 97: ...SmartReader Plus Data Logger 88 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Страница 98: ...SmartReader Plus Data Loggers 89 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Страница 116: ......