SmartReader Plus Data Logger
78
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
4.
Place the logger with sensor or EH-020A module into the jar containing the LiCl solution
(RH = 11%) overnight to eliminate any hysteresis effects.
5.
Quickly insert the logger (or module) into the next solution by swapping the lids.
Transfers should always be made in the direction of higher RH conditions. The jar
should be well sealed. Allow at least two hours to get within three percent of the
equilibrium RH. Repeat for next solution.
Salt
RH
MgCl
2
33%
NaCl
75%
6.
Call up the graph on your computer using TrendReader software. The profile you see
should reflect the test you have just performed.
7.
Zoom in on the section that relates to the 33% RH level in the test chamber. Use the
mouse to find the byte value recorded during the latter part of the minimum two hour test
portion (refer to step 5). Record this as the "bL" reading.
8.
Zoom in on the section that relates to the 75% RH level in the test chamber. Again, use
the mouse to find the byte value recorded at the end of the minimum two hour test (refer
to step 5). Record this value as the "bH" reading.
9.
Switch to the Setup window in TrendReader software.
10. Use the following equation to calculate the "Low" calibration adjustment value:
Low = (71 x b
L
- 145 x b
H
) / 296
11. For the "High" calibration adjustment value use the following equation:
High = 40 x (-74/ (b
H
- b
L
) - 1)
12. Now enter the same values by revising the logger file and view the corrected graphs.
13. If the results are satisfactory, save the calibration changes to the logger (remember to set
the RH channel's equation number back to 71).
Содержание SmartReader Plus
Страница 3: ...SmartReader Plus is a Registered Trademark of ACR Systems Inc TRADEMARK NOTICES ...
Страница 8: ...SmartReader Plus Data Loggers IV Copyright 2010 ACR Systems Inc All Rights Reserved Index 105 ...
Страница 96: ...SmartReader Plus Data Loggers 87 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Страница 97: ...SmartReader Plus Data Logger 88 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Страница 98: ...SmartReader Plus Data Loggers 89 Copyright 2010 ACR Systems Inc All Rights Reserved ...
Страница 116: ......