wire; red and white leads are stripped and tinned.
Figure C-3: ET-081 Probe Resolution
ET-086 Oven Temperature Probe
Description:
NTC thermistor and cable potted into the end of a 22mm
(7/8") long by 5.5mm (7/32") diameter nickel-plated eyelet.
R125 Value:
26,266 ohms
Equation:
48
Range:
70 to 255°C (155 to 490°F)
Max. Temp.:
275°C (525°F)
Resolution:
Refer to Figure C-4
Accuracy:
±1°C (1.8°F) from the temperature values listed in Table C-1
over the range of 70 to 255°C (155 to 490°F)
Cable:
6m (20ft) standard length with twisted-pair moisture-proof
Teflon®- coated glass yarn wire; white leads are stripped and