Setup Environment
Placement:
60 x 60 cm table section
Leave 5 cm of space on the connection
side for plugs and cables
Leave at least 10 cm of space from the
fan for proper ventilation
Leave 5 cm of space from the end of
the table
Leave at least 1 meter of space from the
door for changing of prints and vats
60 cm
60 cm
Leave at least 70 cm of space from the console
for console operations
Place the Diplo on a 60 x 60 cm (2 ft x 2 ft) section of table, leaving 5 cm (2 in) of space on every side
of the machine. Ensure that the fan has at least 10 cm (4 in) of space so air can circulate properly
to avoid overheating.
The front sides – the console side and the door side – both need sufficient working area. Ackuretta
recommends at least 70 cm (2.3 ft) of space to operate the console and at least 1 m (3.3 ft) of space
from the door so you can change prints and vats without obstruction.
UV-Filtered Lighting
The UV-protected door of the Diplo protects the resin from outside lighting while the door is
closed. As you use the printer, the door will be opened many times, and the resin will be exposed to
outside light. To protect the resin, set up the printer in a place where the lighting is controllable or
covered by a UV filter. Similarly, use the same UV filtering in your finishing space and your resin
storage location.
Diplo User Manual
04
Temperature:
15
°
-
35
°
C
•
Keep your Diplo in a dry, temperature-controlled room around 25
°
C (77
°
F).
•
Resins become more viscous when they are too cold. If the environment is around 15
°
C (59
°
F),
Ackuretta recommends increasing all curing times by 15%. Even colder temperatures may require
increasing curing times further or may cause print failures.
•
Similarly, resins may become very thin in hot temperatures. If the environment is hotter than
35
°
C (95
°
F), resins may overcure or may melt during the print process.
Do not operate the Diplo in temperatures above 40
°
C (104
°
F) or below 5
°
C (41
°
F).
Warning: