Photoionization Detector Module PIM CPN
User Manual
PIM_CPN - User Manual (en) - 1.08.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com
Page 23 of 48
Calibration procedure
Action
UART command
I2C data
start
pid.start
Set control register D00
values
pid.values
Read measurement data registers
write
pid.calib
Write calibration data register values
verify
Included in Action ‘write’ with no
error as answer
Set control register D17
Check for status register D17 is set
save
pid.savedata
Set control register D16
Invalid calibration values may be caused by:
•
MEASURE state not started
•
no clean air applied during ZERO calibration step
•
no or inappropriate calibration gas applied during SPAN calibration step
ZERO calibration data
compensated current e.g.
14.000 pA
concentration
0.0 ppm
applying SPAN gas
> 3 min
applying ZERO gas
> 3 min
SPAN calibration data
compensated current e.g.
1334.000 pA
concentration e.g.
100.0 ppm
Action ‘write’
Action ‘verify’
Action 'save’
Action ‘
values
’
Action ‘
values
’
Action ‘start’