Photoionization Detector Module PIM CPN
User Manual
PIM_CPN - User Manual (en) - 1.08.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com
Page 6 of 48
3
Technical data
*
The indicated values were obtained under standardized conditions with 10.6eV lamp.
Test gas was Isobuten in synthetic air.
Detector principle
VUV-Photoionization with 10.6 eV hollow cathode lamp
with Ceramic Discharge Channel technology
and integrated high voltage supply
(optional 11.8 eV and 10.0 eV lamp available)
Detection ranges
0
…
2 000 ppm Isobutene *
Lower detection limit
Typical 10 ppb *
Signal resolution
Dynamic up to 0.01 ppm
Response time
T90 < 10 s *
Influence of humidity
and temperature
compensated between 0 ... 50 °C and 0
…
90 % rH
Operating conditions
-10
…
60 °C
0
…
95 % rH, non-condensing
90 to 110 kPa
Storage conditions
-20
…
60 °C
0
…
95 % rH, non-condensing
90 to 110 kPa
Gas flow
required 200 … 300
ml/min (monitored)
dust and water protection filter at sample inlet required
PID lamp life time
Min. 8 000 hours *, typical more than 15 000 h for 10,6eV lamp
Power supply
3.00 VDC to 5.25 VDC
max. 200 mA peak, typical 90 mA @ 3.30 VDC
Digital interface
UART TTL (at power supply level)
I2C TTL (at power supply level)
Calibration
Two point gas calibration
Dimension, weight
54 mm x 39 mm x 28 mm (L x W x H), about 60 g
Warranty
2 years, except for components in gas path and wearing parts