7.2
Measurement values
Measurement value
Description
Measurement result
Final calibrated, temperature and humidity compensated measurement
result
in ppm units
Current compensated
Temperature and humidity compensated measurement chamber current
in pA units
Temperature
Measurement chamber temperature ( may differ from ambient
temperature )
in °C units
Humidity
Measurement chamber humidity ( may differ from ambient humidity due
to over temperature )
in %rH units
Flow indicator
Measurement chamber gas flow indicator
in % units
7.3
Measurement result resolution
Measurement result
Rounded to resolution
result < 5 ppm
0.01 ppm
result < 10 ppm
0.05 ppm
result < 20 ppm
0.10 ppm
result < 50 ppm
0.50 ppm
result < 120 ppm
1.00 ppm
result < 200 ppm
2.00 ppm
result < 500 ppm
5.00 ppm
result < 1000 ppm
10.00 ppm
result > 1000 ppm
20.00 ppm