Photoionization Detector Module PIM CPN
User Manual
PIM_CPN - User Manual (en) - 1.08.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com
Page 42 of 48
13.1
How To
–
Replace the PID lamp/front isolation/sensor block
Remove cable
1.
Remove the lamp control connector cable from lamp
case.
Remove PID lamp
2.
unscrew the lamp by a slight left twisting motion
Replace front isolation
3.
Remove the old front isolation from the PID sensor block
by using a tweezer, at the notch on front isolation
border.
4.
Place the new front isolation into the PID sensor block
and make sure that the curvature shows to the lamp
side.
Do not bend or touch the front isolation or
measurement chamber with the finger.
Insert PID lamp
5.
Plug the PID lamp with a slight right twisting motion into
the PID sensor block. The last turn will be a little tough
because the O-ring engages.
6.
Connect the lamp control connector cable to the lamp.
7.
Calibrate the device after replace of front isolation or
installation of a new PID lamp.