Photoionization Detector Module PIM CPN
User Manual
PIM_CPN - User Manual (en) - 1.08.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, 12489 Berlin, Germany, www.aci-berlin.com
Page 33 of 48
10.3.4
Module lamp commands
Command
pid.lampinfo ?
pid.lampinfo [serial];[variant];[ontime]
Description
Get the lamp information values
Parameter
serial serial number text of connected lamp
variant variant text of connected lamp
ontime on time of connected lamp in hours units
Example
M
pid.lampinfo ?
S
pid.lampinfo C332003002;106eV;12.500
10.3.5
Module I2C commands
Command
device.address.i2c ?
device.address.i2c [addr]
Description
Get or set the device I2C s
lave
address
Parameter
addr
I2C slace address in decimal ( 1 to 127 )
Example read
M
device.address.i2c ?
S
device.address.i2c 32
Example write
M
device.address.i2c 35
S
device.address.i2c 35
On change the I2C slave address is stored permanently in device.