32
7.8 Calibrating and setting the calibration mode
Electronic balance performs mass measurements using gravity (g). Differences in geographic
regions and altitude vary the gravity acceleration (g).
Therefore, in order to obtain accurate measurements, the balance must be adapted to the
place of use and environmental conditions. This adjustment is performed by the calibration
function.
Select the "
Calibration menu
" function in the settings menu.
In the calibration menu, you can calibrate, display
and print the data for the last calibration and set
the calibration mode.
Before performing the calibration procedure,
check that the plate is empty.
▪
Calibration:
By tapping the calibration key,
the calibration procedure corresponding to the
set calibration mode will be activated.
Load the required calibration weight or tap the
"
Abort
"
button
to
cancel
the
calibration
procedure.
Wait for weight capture.
If the operation is successful, the weighing
screen will display with the calibrated weight
value.
Remove the weight from the plate.